Performance qualification (PQ) is the documented collection of activities necessary to demonstrate that an instrument consistently performs according to the specifications defined by the user, and is appropriate for the intended use. After IQ and OQ have been performed, the instrument's continued suitability for its intended use is demonstrated through performance qualification. The PQ phase may include the following parameters.
Performance Checks— Set up a test or series of tests to verify the acceptable performance of the instrument for its intended use. PQ tests are usually based on the instrument's typical on-site applications and may consist of analyzing known components or standards. The tests should be based on good science and reflect the general intended use of the instrument. Some system suitability tests or quality control checks that are performed concurrently with the test samples can be used to demonstrate that the instrument is performing suitably. PQ tests may resemble those performed during OQ, but the specifications for their results may be set differently if required. Nevertheless, user specifications for PQ tests should demonstrate trouble-free instrument operation for the intended applications. As is the case with OQ testing, PQ tests may be modular or holistic.
Testing frequency depends on the ruggedness of the instrument and the criticality of the tests performed. Testing may be unscheduled—for example, each time the instrument is used. It may also be scheduled for regular intervals. Experience with the instrument can influence this decision. It may be useful to repeat the same PQ tests each time the instrument is used so that a history of the instrument's performance can be compiled. Alternatively, the instrument may be incorporated into an integrated support system to assure that it remains continually qualified. Some system suitability tests or quality control checks that are performed concurrently with the test samples also imply that the instrument is performing suitably.
Preventive Maintenance and Repairs— When an instrument fails to meet PQ test specifications, it requires maintenance or repair. A periodic preventive maintenance may also be recommended for many instruments. The relevant PQ test(s) should be repeated after the needed maintenance or repair to ensure that the instrument remains qualified.
Practices for Operation, Calibration, Maintenance, and Change Control— Establish practices to maintain and calibrate the instrument. Each maintenance and calibration activity should be documented.
Routine analytical tests do not constitute OQ testing. OQ tests are specifically designed to verify the instrument's operation according to specifications in the user's environment, and repeating the testing at regular intervals may not be required. However, when the instrument undergoes major repairs or modifications, relevant OQ and/or PQ tests should be repeated to verify whether the instrument continues to operate satisfactorily. If an instrument is moved to another location, an assessment should be made of what, if any, OQ test should be repeated.
Performance qualification (PQ) is the documented collection of activities necessary to demonstrate that an instrument consistently performs according to the specifications defined by the user, and is appropriate for the intended use. After IQ and OQ have been performed, the instrument's continued suitability for its intended use is demonstrated through performance qualification. The PQ phase may include the following parameters.Performance Checks— Set up a test or series of tests to verify the acceptable performance of the instrument for its intended use. PQ tests are usually based on the instrument's typical on-site applications and may consist of analyzing known components or standards. The tests should be based on good science and reflect the general intended use of the instrument. Some system suitability tests or quality control checks that are performed concurrently with the test samples can be used to demonstrate that the instrument is performing suitably. PQ tests may resemble those performed during OQ, but the specifications for their results may be set differently if required. Nevertheless, user specifications for PQ tests should demonstrate trouble-free instrument operation for the intended applications. As is the case with OQ testing, PQ tests may be modular or holistic.Testing frequency depends on the ruggedness of the instrument and the criticality of the tests performed. Testing may be unscheduled—for example, each time the instrument is used. It may also be scheduled for regular intervals. Experience with the instrument can influence this decision. It may be useful to repeat the same PQ tests each time the instrument is used so that a history of the instrument's performance can be compiled. Alternatively, the instrument may be incorporated into an integrated support system to assure that it remains continually qualified. Some system suitability tests or quality control checks that are performed concurrently with the test samples also imply that the instrument is performing suitably.Preventive Maintenance and Repairs— When an instrument fails to meet PQ test specifications, it requires maintenance or repair. A periodic preventive maintenance may also be recommended for many instruments. The relevant PQ test(s) should be repeated after the needed maintenance or repair to ensure that the instrument remains qualified.Practices for Operation, Calibration, Maintenance, and Change Control— Establish practices to maintain and calibrate the instrument. Each maintenance and calibration activity should be documented.Routine analytical tests do not constitute OQ testing. OQ tests are specifically designed to verify the instrument's operation according to specifications in the user's environment, and repeating the testing at regular intervals may not be required. However, when the instrument undergoes major repairs or modifications, relevant OQ and/or PQ tests should be repeated to verify whether the instrument continues to operate satisfactorily. If an instrument is moved to another location, an assessment should be made of what, if any, OQ test should be repeated.