Analyses were run in sets of six, including four samples, one negative control (deionized water), and one positive control (standard). The average negative control was calculated after six negative controls were run.The negative control point (NCP) which is the cutoff for separating a positive and negative response, was set at 70% of the average NCP, based on the manufacturer’s
recommendations. A new NCP was set for each new lot of reagents.