The requirements of this clause shall apply only to WDs intended to process endoscopes which require a test to verify that the device is watertight.
NOTE This test is intended to demonstrate that the endoscope will not be damaged by liquid ingress during the WD operating cycle. It is regarded only as a test of the integrity of the endoscope when all parameters of the WD leak test (e. g. pressure, duration, maximum leak accepted) are consistent with those specified by the endoscope manufacturer.