Instrument calibration and sample analysis. Silicon was determined in the extracts by AAS,
using a nitrous oxide-acetylene flame. The atomic absorption spectrometer initialization
procedure followed the manufacturer's recommended values (i.e., for gas flows and
lamp current). The 251.6-nm silicon wavelength was used. After the instrument
warm-up period, the gas flow, nebulizer uptake, and burner head alignment adjust-
ments were made while aspirating a Dow Corning © 200 fluid-MIBK solution con-
taining 60 to 70 •xg Si/ml. The final optimization adjustments for best sensitivity and
signal-to-noise ratio were made while aspirating the water-saturated blank solvent and
calibrating solutions.
Calibration in both wet and dry solvent solutions was necessary because silicon is very
sensitive to small changes in flame temperature. The water which saturates the MIBK
during the solvent extraction step alters the temperature and fuel-oxidant stoichiometry
of the N20-C2H2 flame. The atomic absorption response for silicon in wet MIBK is
perhaps reduced to 70% as compared to the same quantity of silicon in dry MIBK.
Dry MIBK silicon calibration solutions were prepared from Dow Corning © 200 fluid,
which has a polymer repeating unit molecular weight of 74. ! g/unit of Me2SiO. It
contains 37.9 percent silicon. Viscosity grades between 100 centistokes and 1000 cen-
tistokes are permissible for use as standards. For a typical standard, 50 mg of DC 200
fluid was weighed into a 250-ml flask partially filled with MIBK. DC 200 fluid tends
to adhere to dry glass surfaces and may require a long contact time with the solvent to
ensure complete solubility of the entire sample. An exact weight of 50 mg is not
essential, but that weight in 250 ml will result in a silicon concentration of 75 }zg of
silicon per milliliter of solvent, or 75 mg/kg Si on a wt/volume basis.
The calibration plot of Si concentration vs. absorbance was linear from the zero inter-
cept through a silicon concentration of 75 mg/kg provided that the atomic absorption
spectrometer was set up under optimum conditions. The exact silicon concentration was
calculated using the following equation:
Sample weight (milligrams + 0.1 mg) x 28.1 x 1000
74.1 x 250ml = micrograms of Si per ml
This weight/volume silicon standard may be serially diluted with MIBK to produce
calibrating solutions lower in silicon content as required.
To make the wet solvent standard calibration solutions, an aliquot of the standard was
shaken with an excess of water, which was centrifuged out, prior to use. For example,
to a 25-ml aliquot of calibrating solution, five ml of water and 0.25 ml of concentrated
HCI reagent were added. Dry standards can be stored and will remain constant, but
wet standards deteriorate on standing and so were made anew for each day's testing. In
addition to the standards, the background solvent which is aspirated into the instru-
ment between samples must also be water-saturated. A supply of MIBK stored over
water was maintained to use for background solvent.
Samples were aspirated into the AA spectrometer, and the absorbance values at 251.6
nm were recorded along with the absorbance values for the wet calibrating solutions. A
calibration plot of absorbance vs. silicon concentration was constructed. The concentra-
tion of silicon in the sample extracts was recorded from the calibration plot, and then
0/5000
จาก: -
เป็น: -
ผลลัพธ์ (ไทย) 1: [สำเนา]
คัดลอก!
Instrument calibration and sample analysis. Silicon was determined in the extracts by AAS, using a nitrous oxide-acetylene flame. The atomic absorption spectrometer initialization procedure followed the manufacturer's recommended values (i.e., for gas flows and lamp current). The 251.6-nm silicon wavelength was used. After the instrument warm-up period, the gas flow, nebulizer uptake, and burner head alignment adjust- ments were made while aspirating a Dow Corning © 200 fluid-MIBK solution con- taining 60 to 70 •xg Si/ml. The final optimization adjustments for best sensitivity and signal-to-noise ratio were made while aspirating the water-saturated blank solvent and calibrating solutions. Calibration in both wet and dry solvent solutions was necessary because silicon is very sensitive to small changes in flame temperature. The water which saturates the MIBK during the solvent extraction step alters the temperature and fuel-oxidant stoichiometry of the N20-C2H2 flame. The atomic absorption response for silicon in wet MIBK is perhaps reduced to 70% as compared to the same quantity of silicon in dry MIBK. Dry MIBK silicon calibration solutions were prepared from Dow Corning © 200 fluid, which has a polymer repeating unit molecular weight of 74. ! g/unit of Me2SiO. It contains 37.9 percent silicon. Viscosity grades between 100 centistokes and 1000 cen- tistokes are permissible for use as standards. For a typical standard, 50 mg of DC 200 fluid was weighed into a 250-ml flask partially filled with MIBK. DC 200 fluid tends
to adhere to dry glass surfaces and may require a long contact time with the solvent to
ensure complete solubility of the entire sample. An exact weight of 50 mg is not
essential, but that weight in 250 ml will result in a silicon concentration of 75 }zg of
silicon per milliliter of solvent, or 75 mg/kg Si on a wt/volume basis.
The calibration plot of Si concentration vs. absorbance was linear from the zero inter-
cept through a silicon concentration of 75 mg/kg provided that the atomic absorption
spectrometer was set up under optimum conditions. The exact silicon concentration was
calculated using the following equation:
Sample weight (milligrams + 0.1 mg) x 28.1 x 1000
74.1 x 250ml = micrograms of Si per ml
This weight/volume silicon standard may be serially diluted with MIBK to produce
calibrating solutions lower in silicon content as required.
To make the wet solvent standard calibration solutions, an aliquot of the standard was
shaken with an excess of water, which was centrifuged out, prior to use. For example,
to a 25-ml aliquot of calibrating solution, five ml of water and 0.25 ml of concentrated
HCI reagent were added. Dry standards can be stored and will remain constant, but
wet standards deteriorate on standing and so were made anew for each day's testing. In
addition to the standards, the background solvent which is aspirated into the instru-
ment between samples must also be water-saturated. A supply of MIBK stored over
water was maintained to use for background solvent.
Samples were aspirated into the AA spectrometer, and the absorbance values at 251.6
nm were recorded along with the absorbance values for the wet calibrating solutions. A
calibration plot of absorbance vs. silicon concentration was constructed. The concentra-
tion of silicon in the sample extracts was recorded from the calibration plot, and then
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 2:[สำเนา]
คัดลอก!
Instrument calibration and sample analysis. Silicon was determined in the extracts by AAS,
using a nitrous oxide-acetylene flame. The atomic absorption spectrometer initialization
procedure followed the manufacturer's recommended values (i.e., for gas flows and
lamp current). The 251.6-nm silicon wavelength was used. After the instrument
warm-up period, the gas flow, nebulizer uptake, and burner head alignment adjust-
ments were made while aspirating a Dow Corning © 200 fluid-MIBK solution con-
taining 60 to 70 •xg Si/ml. The final optimization adjustments for best sensitivity and
signal-to-noise ratio were made while aspirating the water-saturated blank solvent and
calibrating solutions.
Calibration in both wet and dry solvent solutions was necessary because silicon is very
sensitive to small changes in flame temperature. The water which saturates the MIBK
during the solvent extraction step alters the temperature and fuel-oxidant stoichiometry
of the N20-C2H2 flame. The atomic absorption response for silicon in wet MIBK is
perhaps reduced to 70% as compared to the same quantity of silicon in dry MIBK.
Dry MIBK silicon calibration solutions were prepared from Dow Corning © 200 fluid,
which has a polymer repeating unit molecular weight of 74. ! g/unit of Me2SiO. It
contains 37.9 percent silicon. Viscosity grades between 100 centistokes and 1000 cen-
tistokes are permissible for use as standards. For a typical standard, 50 mg of DC 200
fluid was weighed into a 250-ml flask partially filled with MIBK. DC 200 fluid tends
to adhere to dry glass surfaces and may require a long contact time with the solvent to
ensure complete solubility of the entire sample. An exact weight of 50 mg is not
essential, but that weight in 250 ml will result in a silicon concentration of 75 }zg of
silicon per milliliter of solvent, or 75 mg/kg Si on a wt/volume basis.
The calibration plot of Si concentration vs. absorbance was linear from the zero inter-
cept through a silicon concentration of 75 mg/kg provided that the atomic absorption
spectrometer was set up under optimum conditions. The exact silicon concentration was
calculated using the following equation:
Sample weight (milligrams + 0.1 mg) x 28.1 x 1000
74.1 x 250ml = micrograms of Si per ml
This weight/volume silicon standard may be serially diluted with MIBK to produce
calibrating solutions lower in silicon content as required.
To make the wet solvent standard calibration solutions, an aliquot of the standard was
shaken with an excess of water, which was centrifuged out, prior to use. For example,
to a 25-ml aliquot of calibrating solution, five ml of water and 0.25 ml of concentrated
HCI reagent were added. Dry standards can be stored and will remain constant, but
wet standards deteriorate on standing and so were made anew for each day's testing. In
addition to the standards, the background solvent which is aspirated into the instru-
ment between samples must also be water-saturated. A supply of MIBK stored over
water was maintained to use for background solvent.
Samples were aspirated into the AA spectrometer, and the absorbance values at 251.6
nm were recorded along with the absorbance values for the wet calibrating solutions. A
calibration plot of absorbance vs. silicon concentration was constructed. The concentra-
tion of silicon in the sample extracts was recorded from the calibration plot, and then
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 3:[สำเนา]
คัดลอก!
การสอบเทียบเครื่องมือวัดและการวิเคราะห์ตัวอย่าง ซิลิคอน ตัดสินใจใน สารสกัดจาก AAS
โดยใช้เปลวไฟอะเซทิลีนไนตรัสออกไซด์ . การดูดซับอะตอมสเปกเริ่มต้น
ขั้นตอนตามผู้ผลิตแนะนำค่า ( เช่น การเคลื่อนย้ายก๊าซและ
โคมไฟในปัจจุบัน ) การ 251.6-nm ซิลิคอนความยาวคลื่นที่ใช้ หลังจากเครื่องมือ
อุ่นเครื่องระยะเวลา อัตราการไหลของแก๊ส , nebulizer การดูดซึมเครื่องเขียนและการจัดแนวหัวปรับ -
ments เกิดขึ้นขณะที่สำลักเป็นดาว คอร์นนิ่งสงวนลิขสิทธิ์ 200 ของเหลว MIBK โซลูชั่นคอน -
TAINING 60 - 70 - ศรี / ml สามารถเพิ่มประสิทธิภาพการปรับให้ไวที่สุด และสุดท้าย
อัตราส่วนสัญญาณต่อเสียงรบกวนได้ขณะที่สำลักน้ำและตัวทำละลายที่ว่างเปล่าและ
เปิดโซลูชั่น
การสอบเทียบทั้งเปียกและแห้งตัวทำละลาย เป็นเพราะโซลูชั่นซิลิคอนมากไวต่อการเปลี่ยนแปลงอุณหภูมิ
ขนาดเล็กในเปลวไฟ น้ำที่ saturates ที่ MIBK
ในระหว่างขั้นตอนการสกัดด้วยตัวทำละลาย อุณหภูมิ และเชื้อเพลิงเปลี่ยนแปลงอนุมูลอิสระสาร
ของ n20-c2h2 เปลวไฟ การตอบสนองสำหรับการดูดซึมอะตอมซิลิกอน MIBK เปียก
อาจจะลดลงถึง 70% เมื่อเทียบกับปริมาณเดียวกันของซิลิกอนบริการ MIBK .
บริการ MIBK ซิลิคอนปรับแต่งโซลูชั่นเตรียมจาก Dow Corning สงวนลิขสิทธิ์ 200 fluid
ซึ่งมีน้ำหนักโมเลกุลของพอลิเมอร์ย้ำหน่วย 74 ! g / หน่วย me2sio . มัน
ประกอบด้วยซิลิคอน 37.9 เปอร์เซ็นต์ เกรดความหนืดระหว่าง 100 และ 1000 centistokes ศูนย์
tistokes จะอนุญาตให้ใช้เป็นมาตรฐานสำหรับมาตรฐานทั่วไป 50 มก. ของ DC 200
ของเหลวหนักเป็น 250 ml . ขวดบางส่วนที่มี MIBK . DC 200 ของเหลวมีแนวโน้มที่จะยึดติดกับพื้นผิวแก้ว
แห้ง และอาจต้องใช้เวลานานติดต่อกับตัวทำละลายละลาย

ให้สมบูรณ์ของตัวอย่างทั้งหมด น้ำหนักที่แน่นอนของ 50 มก. ไม่
จำเป็น แต่น้ำหนัก 250 มิลลิลิตร จะมีผลในซิลิคอนความเข้มข้น 75 }
ZG ของซิลิคอนต่อมิลลิลิตรของสารละลาย หรือ 75 มก. / กก. ในน้ำหนัก / ปริมาตร ศรีฐาน
การแปลงศรีสมาธิกับการดูดกลืนแสงเป็นเส้นตรงจากศูนย์อินเตอร์ -
แต่ก่อนผ่านซิลิคอนความเข้มข้น 75 มก. / กก. ให้ Atomic absorption
สเปกโตรมิเตอร์ตั้งขึ้นภายใต้สภาวะที่เหมาะสม . ปริมาณซิลิคอนที่แน่นอนคือคำนวณโดยใช้สมการต่อไปนี้ :

น้ำหนักตัวอย่าง ( มิลลิกรัม 0.1 มก. ) x 28.1 x 1000
x 250ml bp = ไมโครกรัมต่อมิลลิลิตรซี
นี้น้ำหนัก / ปริมาตรมาตรฐานซิลิคอนอาจเป็นเจือจางด้วย MIBK ผลิต
เปิดโซลูชั่นลดปริมาณซิลิคอน ตามที่ต้องการ
เพื่อให้เปียกตัวทำละลายสารมาตรฐานโซลูชั่นเป็นส่วนลงตัวของมาตรฐาน
หวั่นไหวกับส่วนเกินของน้ำซึ่งอยู่ที่ระดับออก ก่อนที่จะใช้ตัวอย่างเช่น
เป็นส่วนลงตัวของ 25 ml ปรับโซลูชั่น 5 มิลลิลิตรของน้ำและใช้กรดไฮโดรคลอริกเข้มข้น 0.25 มิลลิลิตร
ถูกเพิ่ม มาตรฐานแห้ง สามารถเก็บไว้และจะคงที่แต่
มาตรฐานเปียกเสื่อมสภาพยืน และก็สร้างใหม่ขึ้นมาสำหรับการทดสอบของแต่ละวัน . ใน
นอกจากมาตรฐาน ฉากหลังละลายซึ่งเป็น aspirated เข้าไปในชุดอุปกรณ์ -
ติดขัดระหว่างตัวอย่างยังต้องน้ำอิ่มตัว อุปทานของ MIBK เก็บไว้มากกว่า
น้ำไว้ใช้เพื่อละลายฉากหลัง
จำนวน aspirated เข้าไปวัดระดับ AA และค่าการดูดกลืนแสงที่ 251.6
nm ที่ถูกบันทึกไว้พร้อมกับค่านเปียกเปิดโซลูชั่น ปรับค่าการดูดกลืนแสงและแปลงเป็น
ซิลิคอนสมาธิถูกสร้างขึ้น ซึ่งครุ่นคิด -
tion ของซิลิคอนในตัวอย่างจากการบันทึกจากพล็อตแล้ว
การแปล กรุณารอสักครู่..
 
ภาษาอื่น ๆ
การสนับสนุนเครื่องมือแปลภาษา: กรีก, กันนาดา, กาลิเชียน, คลิงออน, คอร์สิกา, คาซัค, คาตาลัน, คินยารวันดา, คีร์กิซ, คุชราต, จอร์เจีย, จีน, จีนดั้งเดิม, ชวา, ชิเชวา, ซามัว, ซีบัวโน, ซุนดา, ซูลู, ญี่ปุ่น, ดัตช์, ตรวจหาภาษา, ตุรกี, ทมิฬ, ทาจิก, ทาทาร์, นอร์เวย์, บอสเนีย, บัลแกเรีย, บาสก์, ปัญจาป, ฝรั่งเศส, พาชตู, ฟริเชียน, ฟินแลนด์, ฟิลิปปินส์, ภาษาอินโดนีเซี, มองโกเลีย, มัลทีส, มาซีโดเนีย, มาราฐี, มาลากาซี, มาลายาลัม, มาเลย์, ม้ง, ยิดดิช, ยูเครน, รัสเซีย, ละติน, ลักเซมเบิร์ก, ลัตเวีย, ลาว, ลิทัวเนีย, สวาฮิลี, สวีเดน, สิงหล, สินธี, สเปน, สโลวัก, สโลวีเนีย, อังกฤษ, อัมฮาริก, อาร์เซอร์ไบจัน, อาร์เมเนีย, อาหรับ, อิกโบ, อิตาลี, อุยกูร์, อุสเบกิสถาน, อูรดู, ฮังการี, ฮัวซา, ฮาวาย, ฮินดี, ฮีบรู, เกลิกสกอต, เกาหลี, เขมร, เคิร์ด, เช็ก, เซอร์เบียน, เซโซโท, เดนมาร์ก, เตลูกู, เติร์กเมน, เนปาล, เบงกอล, เบลารุส, เปอร์เซีย, เมารี, เมียนมา (พม่า), เยอรมัน, เวลส์, เวียดนาม, เอสเปอแรนโต, เอสโทเนีย, เฮติครีโอล, แอฟริกา, แอลเบเนีย, โคซา, โครเอเชีย, โชนา, โซมาลี, โปรตุเกส, โปแลนด์, โยรูบา, โรมาเนีย, โอเดีย (โอริยา), ไทย, ไอซ์แลนด์, ไอร์แลนด์, การแปลภาษา.