ProcedureOperating the LI-6400Put in batteries and attach sensor head  การแปล - ProcedureOperating the LI-6400Put in batteries and attach sensor head  ไทย วิธีการพูด

ProcedureOperating the LI-6400Put i

Procedure

Operating the LI-6400

Put in batteries and attach sensor head to console (take care uncoiling the cable to prevent cable wind).
Unwind lid tension screw to tighten cuvette lid on sensor head.
Attach CO2cartridge: check O-ring is present and in good condition. Place cartridge where it lines up with O-ring, then screw on gently until it feels like it meets, then one last hard and fast screw to break the seal on the cartridge, make sure it is all the way on.
Turn on (using on/off switch).
In main menu, go to configuration menu, go to light source control, go to pick source and select the appropriate setting, press enter and escape to main menu.
Go to new measurements, line 2 and select temperature control. Set block temperature to control at your standard temperature.
In main menu, go to calibration menu, go to zero flow meter, you will have to wait about 10 seconds and then after verifying that the voltage is close to zero (otherwise manually adjust) you can hit OK (F5).
A. Humid conditions (where obtaining a valid H2O zero may be difficult without dry N2gas) In the calibration menu, go to IRGA zeroing, make sure the CO2scrub is all the way on and that the desiccant is all the way off (bypass), then you will hit M to turn Match (Mch) on. Stay in this mode for 20 min to allow IRGAs to warm up. Do not hit auto CO2, auto H2O or auto all. The LI-6400 should only be zeroed when running pure N2gas through the system. Verify good zero values and record.
B. Dry conditions (replace used drierite in scrub column with fresh drierite). In the calibration menu, go to IRGA zeroing, turn the H2O scrub on fully, then you will hit M to turn Match (Mch) on. Stay in this mode for 20 min to allow IRGAs to warm up. Verify good zero values and record. Hit auto H2O and then turn the CO2scrub all the way on and the desiccant all the way off (bypass). Wait for a good zero to be reached (2-5 min) then verify good zero values and record. Hit auto CO2, then exit calibration menu.
Go to new measurements menu. Disconnect leaf thermocouple and verify leaf temperature equals block temperature (row h). If not, adjust potentiometer underneath IRGA mounting bracket.
Go to menu (line) 1 and open the log file (F5), name the file (usually use date and species ID - which is first 2 letters of each part of the scientific name, and any other information, like leaf # or plant #, e.g. 011607SOFA1-1) also put a short note, eg. SOFA1-1
Go to menu (line) 2 and set the CO2mixer to reference 400 (menu > CO2mixer > reference > 400) and the lamp to a reasonable value (1500 micromol m-2s-1for plants growing in sunlight, or 500 micromol m-2s-1for shade plants.) Decide on Temp (temperature) control - if using batteries, this consumes power and requires more frequent battery changeover.
View the CO2trace (press ] ) and verify that CO2control has reached the set value. Match the IRGAs (line 1 F5).
Check reference water (H2OR) is acceptable. If too low, add up to 10ml water to the CO2scrub column (soda lime). Set flow rate if lower flow required.
Place the chosen leaf in the cuvette, make sure to cover as much of the cuvette with the leaf as possible. Make sure the leaf is in contact with the thermocouple.
Go to menu (line) 3 and enter leaf area estimate if it does not fully cover the opening. Enter STOMRT (1 for equal density on top and bottom, 0 for stomata on one side only).
Once the cuvette is closed, watch the parameters (Photo, Cond, Ci, Trans). Ensure that transpiration does not lead to excessive humidity. NOTE: Generally humidity greater than 85% is excessive. If condensation occurs, the gas exchange calculations will be wrong until the system is dried out.
View the graphs (press ] ) and when graph looks stable log the point using menu (line) 1, F1, or the button on the sensor head.
Light response curve or A-Ci curve. Go to menu 5, select AUTO PROG, pick type of curve, append to file (Y).

A. Light curve standard protocol: 1500,1000, 500, 250, 120, 60, 40, 20, 10, 0), set min. wait time (usually 120 secs), set max. wait time (usually 200 secs), Y (yes) to match if change in CO2gets less than 20, Y (yes) stability definition is OK.
B. CO2response curve: (for C3species) ambient, 50, 100, 150, 250, 350, 500, 700, 900, 1200, 700, 400. (for C4 species) ambient, 30, 50, 80, 110, 150, 250, 350, 500, 700, 400. set min. wait time (usually 120 secs), set max. wait time (usually 200 secs), Y (yes) to match if change in CO2gets less than 200, Y (yes) stability definition is OK.
While the autoprograms are running, keep an eye on the following:

display line C, Photo (photosynthesis) should be between 0-15 or higher, respiration in darkness 0 to -2, Cond (conductance) should always be positive, between .01- 1, Ci should be positive (a negative value indicates a problem with leaf temperature or water calibration).
display line D, VpdL, if greater than 2, then leaf is too hot, maybe too hot outside or too late in the day to continue (although OK for diurnals).
0/5000
จาก: -
เป็น: -
ผลลัพธ์ (ไทย) 1: [สำเนา]
คัดลอก!
ProcedureOperating the LI-6400Put in batteries and attach sensor head to console (take care uncoiling the cable to prevent cable wind).Unwind lid tension screw to tighten cuvette lid on sensor head.Attach CO2cartridge: check O-ring is present and in good condition. Place cartridge where it lines up with O-ring, then screw on gently until it feels like it meets, then one last hard and fast screw to break the seal on the cartridge, make sure it is all the way on.Turn on (using on/off switch).In main menu, go to configuration menu, go to light source control, go to pick source and select the appropriate setting, press enter and escape to main menu.Go to new measurements, line 2 and select temperature control. Set block temperature to control at your standard temperature.In main menu, go to calibration menu, go to zero flow meter, you will have to wait about 10 seconds and then after verifying that the voltage is close to zero (otherwise manually adjust) you can hit OK (F5).A. Humid conditions (where obtaining a valid H2O zero may be difficult without dry N2gas) In the calibration menu, go to IRGA zeroing, make sure the CO2scrub is all the way on and that the desiccant is all the way off (bypass), then you will hit M to turn Match (Mch) on. Stay in this mode for 20 min to allow IRGAs to warm up. Do not hit auto CO2, auto H2O or auto all. The LI-6400 should only be zeroed when running pure N2gas through the system. Verify good zero values and record.B. แห้งเงื่อนไข (แทน drierite ใช้ในคอลัมน์ขัดกับสด drierite) ในการสอบเทียบ:เมนูไปชีพ IRGA เปิดสครับของ H2O เต็ม จาก นั้นคุณจะกด M เพื่อเปิดการแข่งขัน (Mch) เข้าโหมดนี้สำหรับ 20 นาที IRGAs อุ่นให้ ตรวจสอบดีค่าเป็นศูนย์และบันทึก ตีอัตโนมัติ H2O และเปิด CO2scrub สุด และชื้นให้ปิด (บายพาส) รอศูนย์ดีถึง (2-5 นาที) แล้วตรวจสอบดีค่าเป็นศูนย์และบันทึก ตีอัตโนมัติ CO2 แล้วออกจากเมนูการปรับเทียบไปที่เมนูวัดใหม่ ยกเลิกใบเท และตรวจสอบใบอุณหภูมิเท่ากับอุณหภูมิบล็อก (แถว h) ถ้ายัง ไม่ได้ ปรับมิเตอร์ใต้ IRGA ยึดไปที่เมนู (สาย) 1 และเปิดแฟ้มล็อก (F5) ชื่อแฟ้ม (มักจะใช้วันและพันธุ์ ID - ซึ่งเป็นครั้งแรก 2 ตัวอักษรของแต่ละส่วนของชื่อทางวิทยาศาสตร์ และข้อมูลอื่น ๆ ใบ# หรือ เช่น 011607SOFA1-1 โรงงาน) ยัง ใส่หมายเหตุสั้น ๆ เช่น SOFA1-1ไปที่เมนู (สาย) 2 และตั้งค่า CO2mixer อ้างอิง 400 (เมนู > CO2mixer > อ้างอิง > 400) และโคมไฟเป็นค่าที่เหมาะสม (1500 micromol m-2s-1for พืชเติบโตในแสงแดด หรือพืชสี m-2s-1for micromol 500) ตัดสินใจในการควบคุมอุณหภูมิ (อุณหภูมิ)--ถ้าใช้แบตเตอรี่ นี้ใช้พลังงาน และต้องเปลี่ยนแบตเตอรี่บ่อยดู CO2trace (กด]) และตรวจสอบว่า CO2control เข้าตั้งค่า ตรง IRGAs (สาย 1 F5)Check reference water (H2OR) is acceptable. If too low, add up to 10ml water to the CO2scrub column (soda lime). Set flow rate if lower flow required.Place the chosen leaf in the cuvette, make sure to cover as much of the cuvette with the leaf as possible. Make sure the leaf is in contact with the thermocouple.Go to menu (line) 3 and enter leaf area estimate if it does not fully cover the opening. Enter STOMRT (1 for equal density on top and bottom, 0 for stomata on one side only).Once the cuvette is closed, watch the parameters (Photo, Cond, Ci, Trans). Ensure that transpiration does not lead to excessive humidity. NOTE: Generally humidity greater than 85% is excessive. If condensation occurs, the gas exchange calculations will be wrong until the system is dried out.View the graphs (press ] ) and when graph looks stable log the point using menu (line) 1, F1, or the button on the sensor head.Light response curve or A-Ci curve. Go to menu 5, select AUTO PROG, pick type of curve, append to file (Y).A. Light curve standard protocol: 1500,1000, 500, 250, 120, 60, 40, 20, 10, 0), set min. wait time (usually 120 secs), set max. wait time (usually 200 secs), Y (yes) to match if change in CO2gets less than 20, Y (yes) stability definition is OK.B. CO2response curve: (for C3species) ambient, 50, 100, 150, 250, 350, 500, 700, 900, 1200, 700, 400. (for C4 species) ambient, 30, 50, 80, 110, 150, 250, 350, 500, 700, 400. set min. wait time (usually 120 secs), set max. wait time (usually 200 secs), Y (yes) to match if change in CO2gets less than 200, Y (yes) stability definition is OK.While the autoprograms are running, keep an eye on the following:display line C, Photo (photosynthesis) should be between 0-15 or higher, respiration in darkness 0 to -2, Cond (conductance) should always be positive, between .01- 1, Ci should be positive (a negative value indicates a problem with leaf temperature or water calibration).display line D, VpdL, if greater than 2, then leaf is too hot, maybe too hot outside or too late in the day to continue (although OK for diurnals).
การแปล กรุณารอสักครู่..
ผลลัพธ์ (ไทย) 3:[สำเนา]
คัดลอก!
ขั้นตอนปฏิบัติการ li-6400ใส่แบตเตอรี่และแนบหัวเซ็นเซอร์ที่คอนโซล ( ดูแลสมรรถภาพการวิจัยปฏิบัติการในชั้นเรียนสายเพื่อป้องกันสายลม )ผ่อนคลายความตึงเครียด สกรูขันฝาฝาพิทยาพลบนหัวเซนเซอร์แนบ co2cartridge : ตรวจสอบโอริงเป็นปัจจุบันและสภาพดี ที่ตลับหมึกที่เส้นกับโอริง แล้วขันสกรูบนเบา ๆจนรู้สึกได้ แล้วสุดท้ายอย่างหนักและรวดเร็วสกรูเพื่อทำลายผนึกบนตลับหมึก ให้แน่ใจว่ามันคือทั้งหมดที่เกี่ยวกับเปิด ( ใช้เปิด / ปิด )ในเมนูหลัก ไปที่เมนูการตั้งค่า ไปที่แสงการควบคุมแหล่งที่มาไปเลือกแหล่งที่มาและเลือกการตั้งค่าที่เหมาะสม กด Enter และหลบหนีไปยังเมนูหลักไปที่ วัดใหม่ บรรทัดที่ 2 และเลือกเครื่องควบคุมอุณหภูมิ ชุดป้องกันอุณหภูมิเพื่อควบคุมอุณหภูมิมาตรฐานได้ในเมนูหลัก ไปที่เมนูปรับแต่ง ไปไหลศูนย์เมตรคุณจะต้องรอประมาณ 10 วินาที จากนั้น หลังจากการตรวจสอบแรงดันไฟฟ้าว่าอยู่ใกล้กับศูนย์ ( หรือปรับตนเอง ) คุณสามารถกดปุ่มตกลง ( F5 )A . ชื้น ( ที่ได้รับ H2O ที่ศูนย์อาจจะยากโดย n2gas แห้ง ) ในเมนูปรับแต่ง ไป irga zeroing ให้แน่ใจว่า co2scrub อยู่ตลอดทาง และที่ตัวมีหมด ( บายพาส ) แล้วคุณจะกดปุ่ม M เพื่อเปิดการแข่งขัน ( MCH ) บน อยู่ในโหมดนี้ นาน 20 นาที เพื่อให้ irgas อุ่น ไม่ตี CO2 อัตโนมัติ auto H2O หรือรถยนต์ทั้งหมด การ li-6400 ควรจะจุดต่ำสุดเมื่อวิ่ง n2gas บริสุทธิ์ ผ่านระบบ ตรวจสอบค่าศูนย์ และบันทึกB . แห้งเงื่อนไข ( แทนการใช้สครับ drierite ในคอลัมน์ที่มี drierite สด ) ในเมนูปรับแต่ง ไป irga zeroing เปิด H2O ขัดกับอย่างเต็มที่ แล้วคุณจะกดปุ่ม M เพื่อเปิดการแข่งขัน ( MCH ) บน อยู่ในโหมดนี้ นาน 20 นาที เพื่อให้ irgas อุ่น ตรวจสอบค่าศูนย์ และบันทึก กดปุ่ม Auto H2O แล้วเปิด co2scrub ทุกวิถีทางและสารดูดความชื้นหมด ( บายพาส ) รอดีศูนย์ที่จะถึง ( 2-5 นาที แล้วตรวจสอบค่าศูนย์ และบันทึก กดปุ่ม CO2 อัตโนมัติแล้วออกจากเมนูปรับแต่งไปใหม่ค่าเมนู ใบตัดและใบตรวจสอบอุณหภูมิเท่ากับอุณหภูมิ Thermocouple บล็อก ( แถว H ) ถ้าไม่ปรับ Potentiometer ใต้ irga ยึดติด .ไปที่เมนู ( เส้น ) ที่ 1 และเปิดแฟ้มบันทึก ( F5 ) , ชื่อไฟล์ ( มักจะใช้วันที่และหมายเลข - ซึ่งเป็นชนิดแรก 2 ตัวอักษรของแต่ละส่วนของชื่อทางวิทยาศาสตร์และข้อมูลอื่น ๆเช่น ใบ # หรือ # พืช เช่น 011607sofa1-1 ) ยังเขียนโน้ตสั้นๆ เช่น sofa1-1ไปที่เมนู ( สาย 2 ) และตั้งค่า co2mixer อ้างอิง 400 ( เมนู > co2mixer > การอ้างอิง > 400 ) และโคมไฟกับคุณค่าที่เหมาะสม ( พืช m-2s-1for 1500 micromol เติบโตในแสงแดด หรือ 500 micromol m-2s-1for ร่มเงาพืช ) ตัดสินใจใน Temp ( อุณหภูมิ ) การควบคุม - ถ้าใช้แบตเตอรี่ ซึ่งใช้พลังงาน และต้อง เปลี่ยนแบตเตอรี่ บ่อยกว่าดู co2trace ( กด ) และการตรวจสอบ co2control มาถึงชุดค่า ตรงกับ irgas ( บรรทัดที่ 1 F5 )ตรวจสอบน้ำอ้างอิง ( h2or ) เป็นที่ยอมรับ ถ้าต่ำเกินไป เพิ่มน้ำให้ co2scrub สำหรับคอลัมน์ ( โซดามะนาว ) ตั้งค่าอัตราการไหลถ้าลดกระแสที่ต้องการใบที่เลือกในคิวเวตต์ ให้แน่ใจว่าครอบคลุมมากของคิวเวตต์กับใบมากที่สุด ให้แน่ใจว่าใบไม้มันติดต่อกับเทอร์โมคัปเปิ้ลไปที่เมนู ( สาย ) 3 และระบุพื้นที่ใบ คาดว่าถ้ามันถมไม่เต็มช่อง ระบุ stomrt ( ความหนาแน่นเท่ากับ 1 สำหรับด้านบนและด้านล่าง , 0 สำหรับปากบนด้านใดด้านหนึ่งเท่านั้น )เมื่อคิวเวตต์ปิดดูพารามิเตอร์ ( ภาพถ่าย , คอนโดมิเนียม , CI , trans ) ให้แน่ใจว่าการคายน้ำไม่ก่อให้เกิดความชื้นมากเกินไป หมายเหตุ : โดยทั่วไปความชื้นมากกว่า 85% มีมากเกินไป ถ้าเกิดการควบแน่นเกิดขึ้น การแลกเปลี่ยนก๊าซ การคำนวณจะผิด จนกว่าระบบจะแห้งออกดูที่กราฟ ( กด ) และเมื่อกราฟดูบันทึกจุดโดยใช้เมนูที่มั่นคง ( เส้น ) 1 , F1 , หรือปุ่มบนหัวเซนเซอร์การตอบสนองต่อแสง a-ci โค้งหรือโค้ง ไปที่เมนูเลือกรายการอัตโนมัติ เลือกประเภทของเส้นโค้ง , ผนวกไฟล์ ( Y )A . เส้นโค้งแสงมาตรฐาน : 15001000 , 500 , 250 , 120 , 60 , 40 , 20 , 10 , 0 ) , การตั้งค่านาทีรอเวลา ( ปกติ 120 วินาที ) , การตั้งค่าเวลารอสูงสุด ( ปกติ 200 วินาที ) , y ( ครับ ) ราคาถ้าเปลี่ยน co2gets น้อยกว่า 20 Y ( ครับ ) ความหมาย เสถียรภาพเป็นโอเคB . co2response โค้ง ( c3species ) อุณหภูมิ , 50 , 100 , 150 , 250 , 350 , 500 , 700 , 900 , 1200 , 700 , 400 ( C4 ชนิด ) , 30 , 50 , 80 , 110 , 150 , 250 , 350 , 500 , 700 , 400 ตั้งนาที รอเวลา ( ปกติ 120 วินาที ) , การตั้งค่าเวลารอสูงสุด ( ปกติ 200 วินาที ) , y ( ครับ ) ราคาถ้าเปลี่ยน co2gets น้อยกว่า 200 , Y ( ใช่ ) นิยามความมั่นคงครับในขณะที่ autoprograms กำลังจับตาดูต่อไปนี้ :แสดงสาย C รูป ( แสง ) ควรอยู่ระหว่าง 0-15 หรือสูงกว่าอัตราการหายใจในความมืด 0 - 2 , คอนโดมิเนียม ( ความนำ ) ควรเป็นบวกระหว่างระดับ . 01 - 1 , CI ควรจะบวก ( ค่าติดลบบ่งชี้ว่า มีปัญหากับใบหรือสอบเทียบอุณหภูมิน้ำ )แสดงสาย D vpdl ถ้ามากกว่า 2 แล้วใบไม้ร้อนเกินไป อาจร้อนเกินไปนอกหรือสายในวันต่อไป ( แต่ไม่เป็นไรเพราะ diurnals )
การแปล กรุณารอสักครู่..
 
ภาษาอื่น ๆ
การสนับสนุนเครื่องมือแปลภาษา: กรีก, กันนาดา, กาลิเชียน, คลิงออน, คอร์สิกา, คาซัค, คาตาลัน, คินยารวันดา, คีร์กิซ, คุชราต, จอร์เจีย, จีน, จีนดั้งเดิม, ชวา, ชิเชวา, ซามัว, ซีบัวโน, ซุนดา, ซูลู, ญี่ปุ่น, ดัตช์, ตรวจหาภาษา, ตุรกี, ทมิฬ, ทาจิก, ทาทาร์, นอร์เวย์, บอสเนีย, บัลแกเรีย, บาสก์, ปัญจาป, ฝรั่งเศส, พาชตู, ฟริเชียน, ฟินแลนด์, ฟิลิปปินส์, ภาษาอินโดนีเซี, มองโกเลีย, มัลทีส, มาซีโดเนีย, มาราฐี, มาลากาซี, มาลายาลัม, มาเลย์, ม้ง, ยิดดิช, ยูเครน, รัสเซีย, ละติน, ลักเซมเบิร์ก, ลัตเวีย, ลาว, ลิทัวเนีย, สวาฮิลี, สวีเดน, สิงหล, สินธี, สเปน, สโลวัก, สโลวีเนีย, อังกฤษ, อัมฮาริก, อาร์เซอร์ไบจัน, อาร์เมเนีย, อาหรับ, อิกโบ, อิตาลี, อุยกูร์, อุสเบกิสถาน, อูรดู, ฮังการี, ฮัวซา, ฮาวาย, ฮินดี, ฮีบรู, เกลิกสกอต, เกาหลี, เขมร, เคิร์ด, เช็ก, เซอร์เบียน, เซโซโท, เดนมาร์ก, เตลูกู, เติร์กเมน, เนปาล, เบงกอล, เบลารุส, เปอร์เซีย, เมารี, เมียนมา (พม่า), เยอรมัน, เวลส์, เวียดนาม, เอสเปอแรนโต, เอสโทเนีย, เฮติครีโอล, แอฟริกา, แอลเบเนีย, โคซา, โครเอเชีย, โชนา, โซมาลี, โปรตุเกส, โปแลนด์, โยรูบา, โรมาเนีย, โอเดีย (โอริยา), ไทย, ไอซ์แลนด์, ไอร์แลนด์, การแปลภาษา.

Copyright ©2024 I Love Translation. All reserved.

E-mail: