Finally, we investigated the measurement repeatability of BP
using the 6 models of commercial oscillometric BP devices
from 5 companies and the KRISS simulator. The models were
summarized in Table I. The experimental connection was like
Fig.5 and the KRISS simulator controlled to generate two
conditions; systolic pressure 120 mmHg/diastolic pressure 80
mmHg/pulse 75 and systolic pressure 150 mmHg/diastolic
pressure 94 mmHg/pulse 75. The measurement was
conducted by 5 times for each condition.