The statics of the installation location must correspond to the weight distribution of the system (ภาพที่ 13).
The system must be installed in a sufficiently illuminated area.
The system must not be installed in an area that experiences vibrations.
The floor must have an even surface and be easy to clean.
The floor's surface must be resistant to solvents, impermeable and antistatic.
The system must not be installed in an area that is heated or cooled by local sources of energy such as heaters or air-conditioning systems.
Equipment that causes electromagnetic interference, high-frequency radiation for instance, must not be operated in the vicinity of the system.
The system must be freely accessible from all sides.
The system may only be operated in an electromagnetically controlled environment
The system is in-door used, working in a dry, dust-free enviroment.
The system should be protected against exposure to direct sunlight ,
precipitation and full wind pressure.
The temperature and humidity refer to section 3.5.
.
The rating plate is located at the back of the system, just above the mains connection on the lower left.
It provides the following information:
Manufacturer
Manufacturer's address
Model
Serial number
Year of construction
Line connection
Power Input
Full load current