The determination of lead using ETAAS was performed in a ZEEnit 600
atomic absorption spectrometer (Analytik Jena AG, Jena, Germany)
equipped with a transversely heated pyrolytic graphite coated tubes
with the PIN platform, Zeeman-effect background correction and an
MPE-60 autosampler (Analytik Jena). A lead hollow cathode lamp
(Analytik Jena AG, Jena, Germany) was used as a radiation source with
a current of 4 mA (wavelength 283.3 nm, spectral bandwidth 0.8 nm).
The purge gas employed was argon (purity of 99.997%, White Martins,
Salvador, Brazil) with an internal flow rate of 2.0 L min−1 during all
steps, except during atomization, when the internal flow was stopped.
Analytical signals were measured as integrated absorbance, Aint. The
graphite furnace temperature program optimized is given in Table 2.
The sample volume used in all measurements was 20 μl.
การหาปริมาณตะกั่วที่ใช้แสดงในเรื่อง zeenit 600อะตอมดูดกลืนสเปคโตรมิเตอร์ ( ANALYTIK JENA AG , เยนา , เยอรมนีพร้อมกับอุ่นตามขวางไพโรไลติกแกรไฟต์เคลือบท่อกับ pin แพลตฟอร์ม Zeeman ผลพื้นหลังแก้ไข และmpe-60 autosampler ( ANALYTIK JENA ) นำหลอดไฟหลอดกลวง( ANALYTIK JENA AG , เยนา , เยอรมนี ) ถูกใช้เป็นแหล่งที่มีการแผ่รังสีปัจจุบันของ 4 มา ( ความยาวคลื่น 283.3 nm ความยาวคลื่นแบนด์วิดธ์ 0.8 nm )การกวาดล้างก๊าซอาร์กอน ( ใช้ความบริสุทธิ์ของ 99.997 % สีขาว มาร์ติน ,ซัลวาดอร์ , บราซิล ) กับภายในอัตราการไหล ของ 2.0 ลิตรนาที− 1 ในทั้งหมดขั้นตอน ยกเว้นช่วงที่เป็นละอองเมื่อกระแสภายในก็หยุดสัญญาณวัดค่าการดูดกลืนแสงเชิงบูรณาการไม่ได้ . ที่โปรแกรมเพิ่มประสิทธิภาพให้อุณหภูมิเตาปฏิกรณ์กราไฟท์ในรางที่ 2ตัวอย่างปริมาณที่ใช้ในการวัดทั้งหมด 20 μ L