Graphs and data shown in this panel are explained here below:
On the top left, a V/T graph showing 30 seconds of subject’s breaths. After 30 seconds, the graph is cleared and it
starts graphing from beginning.
On the top center, the Flow/Pbox graph, showing the subject’s current loop and his/her last one. The shown loops
are not stored. These loops are not shown during the SVC or IC tests.
On the top right, two vertical bars (and the correspondent values) for respiratory frequency and box pressure,
updated in real time, useful for monitoring these values.
On the bottom left, the last captured sRaw and the values of all the captured sRaws. It is possible to view another
sRaw by selecting the corresponding value on the list below the graph.
On the bottom center, the last captured TGV and the values of all the captured TGVs. It is possible to view another
TGV by selecting the corresponding value on the list below the graph.
On the bottom right, the last SVC/IC and the values of all the performed SVC/IC. It is possible to view another
SVC/IC manoeuvre by selecting the corresponding value on the list below the graph.
On the very bottom of the screen, a status bar that can show informative messages during the test.