You are trying to design an experiment such that you keep your samples the linear response range of the instrument. You know that as long as the detector current stays below 1 mamp. you will have linear data. You also know that a .100 micromolar solution of lead gives a signal of .00654 mamp. What is the maximum concentration of lead you can analyze for?
What [Pb] gives a 1 mamp signal?