5. Compare the obtained CV with the CV from the manufacturer. If you cannot obtain the
CV similar to the one provided from the manufacturer, polish and clean your CV once
more and do the CV scan again.
6. Repeat steps 1-5 for Pt disk electrode and Au disk electrode. Scan from -0.2 V to
+1.3 V for Pt disk electrode and -0.2 V to +1.6 V for Au disk electrode.