When using nine different concentrations of FITC dye from 0 nM to 2000 nM, the PMT reading shows linear output (R2 = 0.9991) from 600 mV to 925 mV (Figure S2). Thus, the compact fluorescence detector shows sufficient sensitivity for low limit of detection as well as sufficient linearity for quantification.